Bacteria Detection Kit (qPCR Method)

Product Design and Validation References:

  1. Chinese Pharmacopoeia 2020 Edition: 1101 Sterility Test Method
  2. Chinese Pharmacopoeia 2020 Edition: 9201 Pharmaceutical Microbiological Test Alternative Method Validation Guidelines
  3. “Human Stem Cell-derived Pharmaceutical Research and Evaluation Technical Guidelines (Trial)”
  4. “Immune Cell-derived Pharmaceutical Research and Evaluation Technical Guidelines (Trial)”
  5. American Pharmacopoeia USP44: Sterility Tests
  6. Japanese Pharmacopoeia JP17: Sterility Tests
  7. Polymerase Chain Reaction (PCR) Laboratory Inspection Guidelines (2016 Edition)
Bacteria Detection Kit

Product Name

Catalog No.

Product Specification

Storage Conditions

Bacteria Detection Kit (qPCR Method)

BA0001-50

The  test  kit  contains  5×BAC  reaction solution, enzyme, BAC positive control, BAC internal control, PCR dilution buffer, and 50 tests/kit.

Transport and store at -20℃ Shelf life: 12 months.

Details

  • This product is used for the rapid qualitative detection of bacterial contamination in culture media, cell culture materials, and biological products, and test results can be obtained in 2 hours.
  • It can detect five types of bacteria used for sterility testing in the pharmacopoeia: Staphylococcus aureus, Pseudomonas aeruginosa, Bacillus subtilis, Candida albicans, and Escherichia coli. The detection limit is not more than 100 CFU/mL. The detection limits for Pseudomonas aeruginosa and Bacillus subtilis are verified to be 0.5 CFU/mL. The database comparison covers nearly 400 types of bacteria.
  • Using FAM and ROX two-color fluorescent probes, it can detect bacterial target sequences and internal quality control respectively. It has good specificity and can eliminate false-negative results caused by PCR-inhibiting factors in samples. There is no cross-reaction with fungi, mycoplasma, NK cells, and 293 cells and other human derived cells.

Performance Verification I: Specificity

It has been verified that it can detect five kinds of reference bacteria (Staphylococcus aureus, Pseudomonas aeruginosa, Bacillus subtilis, Salmonella, and Escherichia coli) mentioned in the Chinese Pharmacopoeia at a concentration of 100 CFU/mL, and has no cross-reaction with fungi, mycoplasmas, 293 cell DNA, NK cell DNA, and human umbilical cord mesenchymal stem cell DNA.

Test Kit Specificity Verification Results (Detected Bacterial Channel Amplification Ct Values)

Bacteria

Results

Fungi

Results

Mycoplasmas

Results

Cells

Results

Bacillus subtilis

32.04

Aspergillus niger

NoCt/NoCt

Mycoplasma hyorhinis

NoCt/NoCt

293 cells

NoCt/NoCt

Salmonella

25.46

Candida albicans

NoCt/NoCt

Mycoplasma orale

NoCt/NoCt

NK cells

NoCt/NoCt

Escherichia coli

30.46

Candida tropicalis

NoCt/NoCt

Mycoplasma pneumoniae

NoCt/NoCt

Stem cells

NoCt/NoCt

Pseudomonas aeruginosa

34.10

Candida glabrata

NoCt/NoCt

Unidentified mycoplasma

NoCt/NoCt

  

Staphylococcus aureus

35.92

      

Performance Verification II: Detection Limit and Sensitivity

Taking Pseudomonas aeruginosa as an example, using a bacterial liquid concentration of 10-0.05 CFU/mL, the results show that the detection limit of Pseudomonas aeruginosa is 0.5 CFU/mL.

Detection Limit of Pseudomonas aeruginosa

Taking Salmonella as an example, using a bacterial liquid concentration of 5-0.05 CFU/mL, the results show that the detection limit of Salmonella is 0.5 CFU/mL.

Detection Limit of Salmonella

Performance Verification III: Repeatability and Stability Testing

This test kit was repeatedly produced for 3 batches to test the five types of reference bacteria at a concentration of 100 CFU/mL after extraction. The results are shown in the table below. It can stably detect the five types of reference bacteria at 100 CFU/mL, and the results of the three batches are consistent. The test kit has good reproducibility, with a CV < 10%.

Verification Results of the Repeatability and Stability of the Test Kit (Amplification Ct Values)

 

Batch 1

Batch 2

Batch 3

Bacteria

Bacteria

Internal Control

Bacteria

Internal Control

Bacteria

Internal Control

Bacillus subtilis

32.04

30.04

35.19

29.53

35.91

31.79

Salmonella

25.46

29.59

28.01

29.13

26.32

33.80

Escherichia coli

30.46

31.10

30.59

29.56

30.28

31.63

Pseudomonas aeruginosa

34.10

32.00

33.72

29.75

32.11

31.11

Staphylococcus aureus

35.92

31.39

40.55

29.40

38.65

30.59

NTC

NoCt

NoCt

NoCt

NoCt

NoCt

NoCt

NTC

NoCt

NoCt

NoCt

NoCt

NoCt

NoCt

*For other related experimental validations, please refer to the “Technical White Paper” of Yocon Biotechnology’s products.